K77000 Automatic Pour Point Analyzer

Availablity: In stock

Catalog/Part Number: K77000
Brand: KOEHLER
Supplier: KOEHLER
Supplier Location: Holtsville, New York 11742, USA
Inventory Weight: unavailable
Unit of Measurement:
Inventory Brochure: unavailable

Catalog/Part Number:: K77000

Sorry no inventory found for this brand at this time.

Description

Test Method

ASTM D5771 -  For the description of the determination of the cloud point of petroleum products and biodiesel fuels that are transparent in layers 40 mm in thickness, by an automatic instrument using an optical device. This test method covers the range of temperatures from -60 C to +49 C with temperature resolution of 0.1 C.

After insertion of the prescribed test jar containing the specimen into the apparatus, and the initiation of the program, the specimen is cooled incrementally according to the cooling profile listed in Table 1 of the method. The specimen is continuously monitored by a reflective optical system for the formation of a crystalline structure. When the crystallization of the wax in the specimen is detected by the optical system, the temperature is recorded to within 0.1 C resolution. The specimen is then heated to facilitate the start of the next test.

ASTM D5950 - For the determination of pour point of petroleum products by an automatic instrument that tilts the test jar during cooling and detects movement of the surface of the test specimen with an optical device. This test method is designed to cover the range of temperatures from -66 to +51 C.

After preliminary heating, the test specimen is inserted into the automatic pour point apparatus. After starting the program, the specimen is cooled according to the cooling profile listed in the ASTM Method and examined at either 1 C or 3 C intervals. The lowest temperature at which movement of the specimen is detected, by the automatic equipment, is displayed as the pour point.

Features

Fully Automatic

Automatic determination of cloud point by Optical Detection

Automatic determination of Pour Point via the Automatic Tilt Method

Conformance to ASTM D5771 and D5950

Correlates to ASTM D2500, D97, and related specifications

High Performance

Wireless Cloud Point Head

Wireless Pour Point Head

Integrated Cooling System

Automatic Calibration by means of 10 point RTD Calibration, Sample Calibration, and Jacket Calibration

Industrial Touch Screen User Interface

10.4-inch Integrated Color Touch Screen

Software Capabilities, USB & Network Connections

Standard test methods are pre-programmed or user-defined test programs can be created

Clear graph of temperature vs. time for sample and bath temperatures

Results can be easily searched, viewed on screen, printed, and exported to LIMS

4 USB interfaces

Internet (Ethernet) Line

Automated Calibration Features

3 Point Electronics Calibration

10 Point RTD Offset Calibration

Jacket Calibration

Description

The Automatic Cloud and Pour Point Analyzer is a state of the art piece of equipment for measuring cloud point by Optical Detection and pour point with the Automatic Tilt Method. The Cloud Point is an index of the lowest temperature of a petroleum product or a biodiesel fuels utility for certain applications. Wax crystals of sufficient quantity can plug filters used in some fuel systems. The Pour Point of a petroleum product is an index of the lowest temperature of its utility for certain applications. Flow characteristics, such as pour point can be critical for the correct operation of lubricating oil systems, fuel systems, and petroleum blending and pipeline operations.

Catalog/Part Number: K77000 Category:

Technical Specifications

Specifications

Conforms to the specifications of: ASTM D5771, D5950

Excellent Correlation to: ASTM D97, D2500, D5853, D6074, D6158 ; ISO 3015, 3016; IP 15, 219; DIN 51597; FTM 791-201; NF T 60-105; JIS K2269

Detection:

Cloud Point: Optical Detection

Pour Point: Tilting Method

Cooling: Internal Cooling System

Temperature Range: -105 °C to +50 °C (-157 °F to +122 °F)

Temperature Accuracy: ± 0.1 °C

Tilt Interval: 1 °C or 3 °C as per test method (User defined for custom method 1 °C to 5 °C)

Interfaces: USB (4), Ethernet

Display:10.4" Color Touch Screen

Password Protection: Multi-level password capability

Electrical Requirements

220-240V 50/60Hz

Dimensions

WxDxH,in.(cm): 12.5 x 26.25 x 15.5 (31.75x66.68x39.37)

Add a review

You must login to make reviewLogin

  1. Zero(0) Reviews