K90100 Automatic Melting Point Range Apparatus
Catalog/Part Number:: K90100
Description
Test Method
The melting point of a crystalline solid is the temperature at which
the solid to liquid phase transition occurs, referenced at one atmosphere (1
ATM) of pressure.
Features
Conforms to BP Appendix 5 - Method 6 and GLP specifications
Readily interchanged between automatic and manual detection of melting
point ranges
Intelligent Lamp Intensity Control with Soft Start
Storage capacity for up to 20 sample tests
User-interactive software and data entry, including easy alphanumeric
entry of sample name, ID number, and date
User selectable operating modes:
AUTO detection mode: Start/end of melting point range is automatically
detected by a photosensing infrared device. The melting process is recorded and
viewed on-screen in real-time by a CCD camera.
MANUAL detection mode: Start/end of melting point range can be
selected manually with a key-press by user. Sample melting point can be
determined as per BP method by ‘Heat & Cool’ temperature function.
As above, the melting process is recorded and viewed on-screen in
real-time by a CCD camera.
Description
The Automatic Melting Point Apparatus is the latest technology for
microprocessor-based determinations of melting point ranges of crystalline,
powdered and polymeric materials, and is used to assess sample purity. Requires
approximately 5mg of sample spread uniformly on a glass slide, covered with a
glass coverslip. The slide is placed on a uniformly heated, round furnace and
subjected to a heating profile as required by the user. Precise temperature
control gives reproducible results to within 1%. The unit contains an automatic
temperature safety cut-off feature if no melting points are detected 15°C above
the expected melting point or if the oven reaches 315°C. The melting process is
magnified, recorded, and viewed on-screen in real-time by a CCD camera. The
change in physical appearance of the sample with respect to temperature is
recorded, and the start/end of melting is observed automatically. A
representation of the entire process can be printed out in graphical form for
validation.
Technical Specifications
Specifications
Conforms to
the specifications of:
BP Appendix
5-Method 6; GLP
Visual
Image: 10x magnified displayed on monitor
Temperature
Range: ambient + 5 to 315°C
Heating
Rates: 0.2, 0.5, 1.0, 2.0, 3.0, 6.0, 12.0°C/min
Temperature
Readability: 0.1°C
Cooling
Time: 20 minutes (300°C to ambient)
Temperature
Accuracy:
±0.5°C (ambient
+ 5 to 200°C)
±0.8°C (200
to 315°C)
Sample size:
5 mg (approximately)
Sample
Holder:
Glass Slide
≤1mm ±0.02mm thick
Sample
Cover: Glass Coverslip ±0.17mm thick
Temperature
Sensor: Pt-100 (2 wire RTD)
Test
Storage: Up to 20 tests with parameters
Electrical
Requirements
85 to 260
VAC; 48 to 62 Hz
Dimensions
Main Unit:
161⁄2x121⁄4x13 (42x31x33)
Monitor:
8x51⁄2x51⁄2 (20x14x14)
Additional
Accessories
K90100-1
Glass slides (pack of 500)
K90100-2
Cover slips (pack of 1000)
K90100-3
Sampling jig