K87170 Automatic Vacuum Distillation System
Catalog/Part Number:: K87170
Description
Test Method
Determines
the range of boiling points for petroleum products that can be partially or
completely vaporized at a maximum liquid temperature of 400°C at reduced
pressures. The sample is distilled at a controlled, reduced pressure under
conditions that are designed to provide approximately one theoretical plate
fractionation. Initial and final boiling point is measured and a distillation
curve relating volume percent distilled and the atmospheric equivalent boiling
point temperature can be prepared.
Features
Fully Automatic Operation
Simple to Operate and Maintain
Vacuum Step Down Inhibits Foaming
Easy Access to all Components
Safety Shields and Doors protect Operator
Turn Key System
High Precision and Accuracy
Automatic Cleaning Cycle
Receiver is Easy to Remove
PC Control
Description
The
Automatic Vacuum Distillation System is designed to make vacuum distillation
easy, safe and affordable. The self contained unit is controlled by a standard
PC. Fully automatic function minimizes the amount of operator time needed for
the test. The latest Windows® operating system is included along with a state
of the art PC. The Windows®-based software is intuitive and guides you through
the distillation step by step. All data is saved to the hard drive in standard
format that can be easily opened by spread sheets or exported to LIMS. Files
can be accessed through portable USB drives, Ethernet connection or written to
a CD/DVD. A color printer is provided to print hard copies of the reports.
Process diagrams clearly show the current equipment status. Results can be
viewed as the distillation proceeds in both tables and graphs. Distillation
parameters can be modified at any time during the distillation.
Included
Accessories
Distillation
Column
Mounting
Rack
500mL
Boiling Flask
Electric
Heating Mantle
Lab Jack
Boiler
Temperature Probe
Vapor
Temperature Probe
200mL
Receiver
Clamps
Boiling
Chops
Bath for
heating of condenser, side arm, and receiver
Automatic
Vacuum System
PC for
Distillation Control
Technical Specifications
Specifications
Conforms to
the Specifications of: ASTM D1160; ISO 6616
Condenser
Temperature Range: Ambient +5°C to 150°C
Vacuum
Range: 1.00 mmHg to 50 mmHg (0.13 to 6.7 kPa)
Distillation
Temperature Range: Ambient to 400°C (752°F)
Dimensions
(HxWxD):
91.5 x 94 x 53 cm (36 x 37 x 21 in)
Zero(0) Reviews