Automatic Abel Flash Point Tester
Catalog/Part Number:: K87300
Description
Test Method
For flash point determination of petroleum
products and other liquids having flash points between -30°C and 110°C.
Features
·
Conforms
to IP 170 and related specifications
·
Simple
automation routine for easy operation
·
Data
Storage 200 results
·
Quick
Access to calibration parameters
·
Auto
Diagnostic
·
Optional
low gas cartridge to wok in all types of environments
·
Automatic
gas cut-odd at the end of the test
·
Optional
cryostat required for samples having a FP below 40°C
·
Automatic
Electronic Control of the heating rate during testing
·
Heating by
Solid Block Element
·
Cooling by
External Cryostat
Description
The automated Abel flash point tester is
used primarily to test flammable and combustible materials for shipping and
safety regulations. The flash tester provides an increased temperature range of
operation as compared with other testers, allowing greater flexibility in
testing samples according to the Abel test method. The unit provides a test range
to 110°C and can be extended to –30°C by any appropriate external chiller. The
flash point tests are simply conducted by mounting the flash cup filled with
sample into the test position and selecting a pre- or user-programmed test
method. Automation routines provide accurate test results. A quick search
method is available to determine the flash point of unknown samples. The dual
detection system (thermal and ionization) allows for testing all types of
products. Ignition by gas flame or electrical ignitor is included, along with
safety cut-off devices. Test results are automatically corrected to standard
pressure (101.3 kPa). The system is equipped with a differential Pt-100 RTD
probe designed to duplicate the response time of a mercury-in-glass thermometer
and with multiple sensors that continually monitor instrument function,
displaying an error message if a problem is detected. Supervision software is
included.
Included Accessories
SEIKO ticket printer
Abel Cup
Abel Cover
Pt 100 Probe
Detection Cable
Flexible Stirrer Shaft and Motor
Insulated Tubing for Cryostat
connection
Supervisor Software
Technical Specifications
Specifications
Conforms to the specifications
of:
IP 170; ISO 13016; NF T 66-009;
DIN 51755
Temperature Range: Ambient to
110°C; Subambient to -40°C with external cooling.
Electrical
Requirements
115V
50/60Hz, Single Phase
230V
50/60Hz, Single Phase
Dimensions
lxwxh,in.(cm):
10.25x21x19.75 (26x5.53x50)
Net
Weight: 60 lbs (27.2 kg)
Zero(0) Reviews